On 4/5/07, Brooke Clarke wrote: > Hi Vasile: > > The referenced module is specified for the range of -27 to 482=B0F (-33 to > 220=B0C). So Accuracy =B12% of reading or =B12=B0C means that for temper= atures of 100c > and lower the accuracy is +/- 2%, above 100C it's capped at +/-2 deg C. When the ambiant temperature where device is running is 25C... and the emisivity of the target is selected appropiately. With a wrong emisivity, the error may be up to 20% or more. I have really doubts that module it can measure in -25C...+30C with +/-2% acuracy, but that is my problem... For a real temperature measurement a ratiometric infrared tehnique could be the true solution. -- = http://www.piclist.com PIC/SX FAQ & list archive View/change your membership options at http://mailman.mit.edu/mailman/listinfo/piclist